Checking The Accuracy Of The Tube Voltage; Reproducibility; Measurement Set-Up - Siemens ARCADIS Avantic Maintenance Instruction

Hide thumbs Also See for ARCADIS Avantic:
Table of Contents

Advertisement

Function tests

1 Checking the accuracy of the tube voltage

The actual kV value must agree with the indicated value within the limits stated by
Siemens (max. deviation
nected to a power supply corresponding to Siemens specifications.
Tools: PTW-Nomex
Test procedure:
Refer to the information on the Nomex in the operating instructions
Connect the Nomex.
Remove the single-tank cap at the POWERPHOS after removing the two screws.
Select full image intensifier format and open the collimator fully.
Place the measuring detector on the diaphragm.
Set the following values in the operating field and check
Operating mode
Fluoroscopy
Fluoroscopy
Fluoroscopy

2 Reproducibility

The coefficient of variation must not exceed 0.05 when the X-ray system is connected to a
power supply as specified by Siemens.
The technique factors should be adjusted to alternate settings and reset to the test setting
after each measurement.

2.1 Measurement set-up

Place the dosimeter in the center of the X-ray beam.
Before each new dose measurement, make sure that the indicator is reset to zero.
Siemens AG
Medical Solutions
&
10 % for the ARCADIS Avantic), if the X-ray system is con-
Indicated
70 kV
112 kV
125 kV
SPR2-330.662.01
Rev. 02
05.06
Measured
101 kV to 123 kV
113 kV to 137 kV
Page 1 of 6
CS PS SP
5 - 1
5
5
Tolerance
63 kV to 77 kV
5
5
ARCADIS Avantic