Hydrim

For best results:

Use Stat-Dri Plus rinse aid (part number 8OZPLUS) to prevent spotting.
P0 should be used to keep instruments moist when they will not be washed within one hour.
P1 and P2 should be used to wash instruments in baskets.
P3 and P4 should be used to wash instruments in cassettes and heavily soiled instruments.
Caution:
1.
Hydrim Cleaning Solution is compatible with solid
stainless steel instruments that are resistant to
heat and alkalinity.
2.
Carbon steel, aluminum, nickel and rhodium-plated
instruments may corrode in Hydrim Cleaning Solution.
Do not process burrs or endofiles in the Hydrim.
3.
Do not process stainless steel and carbon steel
instruments at the same time.

Replacing the Hydrim Cleaning Solution:

When the chemical indicator light shines continuously, and Ed appears on the display, the solution level
is low and the container is in need of replacement.
1.
Open the cleaning solution door.
2.
Put a new cleaning solution box into the drawer upside down.
3.
Tear open the tab.
4a.
Pull out the spout and remove the protective cap.
4b.
Attach the hose tightly to the connector on the cleaning solution box.
5.
Place the box in the cleaning solution drawer and then close.
1
®
L110w
Quick Reference Guide
2
3
Shines when cleaning solution needs changing
Flashes when rinse-aid needs filling
Shines when salt needs refilling
*Actual cycle time will vary depending on
incoming water temperature and pressure
4a
4b
8 min.*
20 min.*
30 min.*
30 min.*
40 min.*
5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Hydrim L11w and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SciCan Hydrim L11w

  • Page 1 Hydrim ® L110w Quick Reference Guide For best results: Use Stat-Dri Plus rinse aid (part number 8OZPLUS) to prevent spotting. P0 should be used to keep instruments moist when they will not be washed within one hour. P1 and P2 should be used to wash instruments in baskets. P3 and P4 should be used to wash instruments in cassettes and heavily soiled instruments.
  • Page 2 (maximum of 1kg / 2.2 lbs.) Close the salt container lid. For all service and repair inquiries: United States: 1-800-572-1211, Canada: 1-800-870-7777, International: (416) 446-4500 Email: techservice.ca@scican.com Hydrim is a registered trademark of SciCan, Division of Lux & Zwingenberger Ltd.

Table of Contents