Agilent Technologies E5080 Series User Manual page 23

Cryogen monitor with pressure control
Table of Contents

Advertisement

Problem
Burst disc, burst.
Disconnected burst
disc or open circuit
cable
Non cycling gas
heater
Cryo-cooler low
efficiency
E5080 Series Cryogen Monitor User Guide
As detected by E5083
Burst disc sensor
detected as burst
As above
Gas heater fixed on or off
for longer than about 2
hours, and has not cycled
on - off in that time
A cooler on to off ratio
that exceeds the set
alarm level
Alarm result
Comments
Pressure will fall
Magnet may have quenched
Likely that a
pressure error is
triggered
May be a helium
Can occur post helium fill,
leak if the heater is
prior to re-pressurizing or
on for a long period.
during installation phase
May be lack of
cooling efficiency if
the heater is off for a
long period
Alarm LED active,
sounder active, if not
muted.
2
Operation
21

Advertisement