Connector for power cable
A.
2.2.3.4
Filling and priming procedure
The ICS AirCal uses water to cool the air. It is required that the water reservoir be filled prior to use of the device.
Warning
To prevent electrical shock, the ICS AirCal irrigator must be turned off and disconnected from the electrical
wall outlet when filling or refilling the water reservoir.
Caution
Use this procedure if: 1) the ICS AirCal was just received from shipping dry. It needs to be filled and primed,
2) If the water level in the display has gone to zero because user never refilled it when orange bar was present. 3) If the
ICS AirCal was tilted at more than 20° with the water pump running. If none of these criteria were met and the water
level display is showing an orange bar and needing refill, please follow the instructions in
Caution
Only
the ICS AirCal.
Filling
1. Turn off the ICS AirCal irrigator and unplug from the electrical wall outlet.
2. Wipe off any dust on the top of the case.
3. Facing the front panel, locate the small cover at the back right side. Using the Allen key (supplied with the cradle)
unscrew the reservoir cover and remove.
Natus Medical – ICS® AirCal
use distilled water. Using common water from the tap will result in mineral deposits and will harm
2 Getting started
Refill Procedure: ► 15
9