2 Getting started
13. Remove the funnel and replace the internal cover. Tighten the screws to secure.
14. Replace the external cover and tighten screws to secure.
15. Plug in the irrigator into the electrical wall outlet and turn on.
Note
The default airflow rate for caloric irrigation is 8 liters per minute. Shipping may cause the airflow knob to
•
move. You may have to reset the airflow.
Pr i m in g with bent tube
1. Take the syringe with bent tube and remove enough water from the reservoir to fill the syringe.
2. Put the end of the bent tube into the pump inlet hole. The tube will be facing toward the front of the irrigator.
3. While pressing the priming switch
4. Release the priming switch.
5. Press the priming switch to turn on and hold 5 seconds.
6. Release the priming switch to turn off and wait 5 seconds.
7. Repeat steps
8. Press the cool button
9. If the temperature reaches the set temperature, then the irrigator is primed. Continue to step
10. If the temperature does not reach 12°C in approximately 3 minutes or,repeat steps
11. Turn off the irrigator and unplug from the electrical wall outlet.
12. Check if the water reservoir is full. If not, place the funnel in the refill hole of the reservoir and add additional water.
13. Remove the funnel and replace the internal cover. Tighten the screws to secure.
14. Replace the external cover and tighten screws to secure.
15. Plug in the irrigator into the electrical wall outlet and turn on.
12
, inject the water into the inlet.
5
6
and
for a total of 10 times on and 10 times off.
. Decrease the temperature to 12°C.
11
.
1
9
through
.
Natus Medical – ICS® AirCal
]