A.
Funnel
6. Pour water into the funnel to fill the reservoir.
7. Monitor the water level closely. Fill to the bottom of the lip. Do NOT overfill the reservoir.
8. Remove the funnel.
9. Wipe up any excess water that might have spilled into the small indentions where the internal cover is placed.
10. Replace the internal cover.
Priming
1. Plug in the irrigator into the electrical wall outlet and turn on.
2. Turn off the airflow.
3. Locate the priming switch on the back panel
4. Press the priming switch to turn on and hold 5 seconds.
5. Release the priming switch to turn off and wait 5 seconds.
6. Repeat steps
7. Press the cool button
8. If the temperature reaches the set temperature, then the irrigator is primed. Continue to step
9. If the temperature does not reach 12°C in approximately 3 minutes or, if the over temperature alarm sounds, skip to
the procedure
10. Turn off the irrigator and unplug from the electrical wall outlet.
11. Remove the internal cover.
12. Check if the water reservoir is full. If not, place the funnel in the refill hole of the reservoir and add additional water.
Natus Medical – ICS® AirCal
]
4
5
and
for a total of 10 times on and 10 times off.
. Decrease the temperature to 12°C.
Priming with bent tube ► 12
.
.