10
Guidance and manufacturer's declaration tables
Electromagnetic Emissions
The ICS AirCal device is intended for use in the electromagnetic environment specified below. The customer or user of
the ICS AirCal device should ensure that it is used in such an environment.
Emissions test
Radio Frequency (RF) Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Electromagnetic Immunity - All Equipment and Systems
The ICS AirCal system is intended for use in the electromagnetic environment specified below. The customer or the user
of the system should assure that it is used in such an environment.
I mm unity test
ESD IEC 61000-4-2
EFT IEC 61000-4-4
Surge IEC 61000-4-5
Natus Medical – ICS® AirCal
Compliance
Group 1
Class A
Class A
Complies
IEC 60601 test level
Compliance level
±6kV Contact
±6kV Contact
±8kV Air
±8kV Air
±2kV Mains
±2kV Mains
±1kV Differential
±1kV Differential
±2kV Common
±2kV Common
10 Guidance and manufacturer's declaration tables
Electromagnetic environment - Guidance
The ICS AirCal device uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equip-
ment.
The ICS AirCal device is suitable for use in all establishments
other than domestic and those directly connected to the pub-
lic low-voltage power supply network that supplies buildings
used for domestic purposes.
Electromagnetic environment -
Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a typ-
ical commercial or hospital environment.
Mains power quality should be that of a typ-
ical commercial or hospital environment.