Halo Photonics Stream Line Pro Hardware And Software Manual page 11

Scanning doppler lidar system
Table of Contents

Advertisement

The arrow on the power connector faces
away from the Lidar when it is the correct
orientation. The connector is a ¼ turn
bayonet type. Ensure that it is properly
inserted and locked into position. The
connector may be plastic or metal type.
The metal type still has an orientation
arrow, but it won't be painted. Ensure that
the connector locks into place when
interfacing. The connector may need to be
supported during the connecting process.
This picture shows the green GPS
connector, one spare mast connector
another un-wired smaller connector, and
the white purging inlet.
The purge inlet connector will be airtight as
long as the mating input connector is not
installed.
A corresponding enclosure retaining screw labelled
'Vent' needs to be removed if a dry air purge of the
system is required (i.e. if the internal humidity of the
system becomes high >60%).
A suitable connector for use with the purge inlet can be
seen here:
http://uk.rs-online.com/web/p/hose-
couplings/3872640/
Please contact Halo for advice regarding the purging
procedure if needed.

Advertisement

Table of Contents
loading

Related Products for Halo Photonics Stream Line Pro

Table of Contents