Proline Promass 40
3.3
Post-connection check
• Are cables or the device damaged (visual inspection)?
• Does the supply voltage match the specifications on the nameplate?
• Are the power supply and signal cables correctly connected?
• Do the cables used comply with the necessary specifications?
• Do the mounted cables have adequate strain relief and are they routed securely?
• Is the cable type route completely isolated? Without loops and crossovers?
• Are all screw terminals correctly tightened?
• Are all the cable entries installed, correctly tightened and properly sealed?
• Cable routed as a "water trap" in loops?
• Are all housing covers installed and correctly tightened?
Endress+Hauser
Wiring
11