Apr Adapter - MSA G1 Operating Manual

Air-purifying respirator full facepiece respirator
Hide thumbs Also See for G1:
Table of Contents

Advertisement

US
7. Let the facepiece air-dry. Do not put parts near a heater or in direct sunlight to dry.
8. Operate the exhalation valve manually to make sure that it operates correctly.
9. Do a negative pressure seal test before returning the facepiece to service.
10.2

APR Adapter

WARNING!
• Do NOT use cleaning products that contain hydrocarbons or solvents such as nitro-thinner.
• Do NOT use radiant heat such as the sun or radiators to dry cleaned parts.
• When a drying cabinet is used, make sure that the temperature is not more than 140°F (60°C).
• Do a negative pressure seal test after every cleaning, disinfecting, and maintenance procedure, and after every
exchange of parts.
Failure to follow these warnings can result in serious personal injury or death.
10.2.1 Single Port
1. Use the instructions on the label to prepare a solution of Confidence Plus Germicidal Cleaner (PN 10009971) in a
bucket or sink.
29
5. Rinse the facepiece and components thoroughly in clean water that is not more
than 110°F (43°C) and is preferably running and draining.
6. To clean and rinse the pressure-demand exhalation valve, use a blunt object to
push in on the stem and flush the valve with clean water.
2. Remove the adapter cover.
G1 APR Facepiece
10 Cleaning and Disinfecting