Unit

To change the unit used in MC5-IS' display, press
D
B
8
4
Fieldbus and MC5-IS' Basic Mode
Menu
/
,
Window 2 Setup
/
(opens the second page of the menu) and
Display Unit
/
.
Notes.
In Basic Mode, the unit used in MC5-IS' display need not
be the same as in the instrument itself. Fieldbus instru-
ments use numeric codes for the units they support. When
connecting to an instrument, MC5-IS gets the unit from the
instrument. If the unit is not supported in MC5-IS, the unit's
numeric code is shown and the Quantity is set to "None".
Verifying Calibration and Trimming is not possible when a
unit not supported by MC5-IS is in use in the fieldbus in-
strument. See Appendix 1, Codes in Profibus PA Profile
Version 3.0 on page 34 or Appendix 2, Codes in Profi-
bus PA Profile Version 2.0 on page 39, for lists of sup-
ported units.
This setting does not alter the unit used in the instrument.
For information on how to change the instrument's unit,
see chapter Transducer Block Information on page 23.
User Guide – MC5-IS Fieldbus Option for PROFIBUS PA
, if needed
27