User Guide – MC5-IS Fieldbus Option for PROFIBUS PA
Target Mode
Target Mode is an important Transducer Block parameter
while calibrating instruments. That's why it is directly available
via MC5-IS' menu.
Use this to set the instrument in either OOS - Out Of Service
or Auto - Automatic mode.
See also chapter Notes and Warnings on page 8.
Select Output
Select Output is another important Transducer Block parame-
ter directly available via MC5-IS' menu. The available options
vary depending on the instrument at hand.
Some instruments require that another output is selected when
trimming/adjusting the instrument. That is why the selection of
output is made easily available.
Trim Fieldbus Device
This option allows you to trim/adjust a fieldbus instrument
without performing a Verifying Calibration.
When selected, you first need to enter instrument data as pre-
sented in MC5-IS User Guide's Section D (Chapter Maintain-
ing MC5's Instrument database).
The rest of the trimming/adjustment procedure is done as pre-
sented in chapter Calibrating (Trimming, Adjusting) a Field-
bus Instrument on page 30 of this manual.
24
Fieldbus and MC5-IS' Basic Mode