User Guide – MC5-IS Fieldbus Option for PROFIBUS PA
Performing a Verifying
Calibration

Selecting the Instrument

To be able to calibrate a fieldbus instrument, it has to be added
to MC5-IS' instrument database.
Adding the instruments can be done using the utility described
in chapter Add a Fieldbus Device to Database on page 25.
Additionally, you can also manually enter the instrument data
into MC5-IS' instrument database.
From Basic Mode, go to Calibration Mode (
Mode
IS' list of available instruments.
If you are already connected to the fieldbus instrument (field-
bus communication started in Basic Mode), MC5-IS continues
directly from the Instrument Window to the Calibration Win-
dows.
If fieldbus communication
is not started, MC5-IS
prompts you to start the
communication and se-
lect the instrument in a
window similar to the ad-
jacent picture.
MC5-IS accepts the
fieldbus instrument as
the instrument to be cali-
brated only if its instru-
ment data matches with
the data of the instru-
ment selected in MC5-IS.
Note.
The instrument data in
MC5-IS and in the instrument to be connected need to match.
28
). Then select the instrument to be calibrated from MC5-
3 0 . 0 3 . 2 0 0 7
1 2 : 4 0
P R O F I B U S P A
P A - 1 . 0 0 . 0 . 0 0 . r e l
D e v i c e s f o u n d ( N o d e A d d r : T a g ) :
0 1 6 : T T 1 2 3 4 - P A
B a c k
R e s t a r t
E d i t

Performing a Verifying Calibration