Batteries; Charging The Batteries - BEAMEX MC5-IS User Manual

Fieldbus option for profibus pa
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Batteries

The rechargeable battery inside FI5-IS provides supply power
to the communication between MC5-IS and FI5-IS.
With fresh, fully charged batteries, FI5-IS may be used for ap-
prox. 10 hours without recharging. To save batteries, discon-
nect the fieldbus cables when FI5-IS is not used. When the
front panel light is not visible, there is a minimal load on the
batteries.
Use the battery charger, BC09-IS for charging the batteries.
Using other chargers may put the safety at risk.

Charging the Batteries

The charger, BC09-IS, for rechargeable batteries operates in
the following environments:
The charging time is 16 hours. When taking FI5-IS into use for
the first time, fully charge FI5-IS batteries (16 hours).
Warnings!
USE CHARGER IN SAFE AREA ONLY. The casing of the
charger is not IS-compliant.
USE ONLY THE CHARGER PROVIDED WITH FI5-IS.
The charger should only be used indoors and when charging,
the environment temperature should be between 0 ... +35 °C
(32 ... 95 °F).
See also General Warnings on page 8.
Front Panel Light
The light on the front panel of FI5-IS is red while charging.
When not charging, fieldbus cables are connected and fieldbus
power for FI5-IS is available, the light is green.
If FI5-IS battery is almost empty, the green light is dimmed. If
battery is completely discharged, the green light is not visible.
General
User Guide – MC5-IS Fieldbus Option for PROFIBUS PA
- Voltage: 100 ... 240 VAC
- Frequency:
Note.
Although the led is lit (green), communication is not in-
itiated until MC5-IS is connected to FI5-IS.
50/60 Hz
3