Doing the Verifying Calibration

Doing a Verifying Cali-
bration for a fieldbus in-
strument does not differ
from the calibration of a
non-fieldbus instrument
with similar input quanti-
ty, input method and out-
put quantity. Refer to the
examples in MC5-IS Us-
er Guide's Part D.
Performing a Verifying Calibration
User Guide – MC5-IS Fieldbus Option for PROFIBUS PA
Note.
When adding a fieldbus instrument to MC5-IS, the mea-
surement unit is fetched from the fieldbus instrument. If
MC5-IS does not support the unit, you need to change it to
a unit supported by MC5-IS. Verifying Calibration and
Trimming is not possible when a unit not supported by
MC5-IS is in use in the fieldbus instrument. See Appendix
1, Codes in Profibus PA Profile Version 3.0 on page 34
or Appendix 2, Codes in Profibus PA Profile Version
2.0 on page 39, for lists of supported units.
For information on how to change the instrument's unit,
see chapter Transducer Block Information on page 23.
3 0 . 0 3 . 2 0 0 7
1 5 : 0 6
I n p u t
R T D T e m p e r a t u r e [ E T : S i m u l ]
7 5 . 0 0
O u t p u t
T e m p e r a t u r e [ F I 5 - P A / P V ]
7 5 . 1 6
S t a t u s : G o o d ( N o n - c a s c a d e )
E r r o r
1 . 0 0
0 . 1 6 %
+
0
-
0 %
5 0 %
F o r c e
P a u s e
A c c e p t
P t 1 0 0 a 3 8 5
° C ( I T S 9 0 )
° C
o f s p a n
1 0 0 %