Emerson MICRO MOTION HFVM Series Configuration And Use Manual page 48

Heavy fuel viscosity meters
Table of Contents

Advertisement

Configure process measurement
5.
Option
Description
Internal RTD tem-
Temperature data from the on-
perature data
board temperature sensor
(RTD) is used.
Polling
The meter polls an external de-
vice for temperature data. This
data will be available in addi-
tion to the internal RTD tem-
perature data.
Digital communica-
A host writes temperature data
tions
to the meter at appropriate in-
tervals. This data will be availa-
ble in addition to the internal
RTD temperature data.
6.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed
in the Inputs group on the ProLink III main window.
Need help?
42
Choose the method to be used to supply temperature data, and perform the
required setup.
Verify that your temperature setup is being applied as intended.
a. Choose Device Tools > Configuration > I/O > Inputs > External Inputs.
b. Check or uncheck the checkboxes as desired.
If a checkbox is checked, the internal RTD temperature is used for that
measurement or calculation. If a checkbox is unchecked, the external
temperature is used.
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
-
Verify the wiring between the meter and the external device.
Micro Motion
Setup
a. Set Line Temperature Source to Internal RTD.
b. Click Apply.
a. Set Line Temperature Source to Poll for External Value.
b. Set Polling Slot to an available slot.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option
Description
Poll as Primary
No other HART masters will be on the
network. The Field Communicator is not
a HART master.
Poll as Secondary
Other HART masters will be on the net-
work. The Field Communicator is not a
HART master.
d. Set External Device Tag to the HART tag of the temperature
device.
e. Click Apply.
a. Set Line Temperature Source to Fixed Value or Digital Communica-
tions.
b. Click Apply.
c. Perform the necessary host programming and communica-
tions setup to write temperature data to the meter at appro-
priate intervals.
®