5.3.3
Replacing the fuses
889 IC Sample Center Basic
4 Remove the air needle.
5 Tighten the height adjustment nut on the fixing nut (made of
chrome). The thread of the height adjustment nut must match the
lower part of the fixing nut. The O-ring must be properly seated in
the fixing nut.
6 Insert the air needle.
7 Mount the sample needle.
8 Set the correct needle height for the new needle in the manual con-
trol feature of MagIC Net on the Needle tab.
NOTICE
If you are using the sample racks with 12 or 48 vials, then you
must adjust the setting of the needle height to > 2 mm. This will
prevent the needle from touching the bottom of the vial.
9 Clean the new needle.
The fuses installed in the 889 IC Sample Center Basic are of the type 2 ×
2.5 A.
WARNING
The device must be disconnected from the electricity source before the
fuses are replaced.
Make sure that the fuses to be installed correspond to the correct type
and the correct power load.
The fuses are located in the fuse holder at the rear of the device.
NOTICE
Contact a regional Metrohm service representative in the event of
repeated problems with the fuses.