Akkulift ® – PRODUCT DESCRIPTION • If Akkulift has been exposed to low temperatures ® for a longer period of time (e.g. longer storage in Many thanks for deciding for AKKULIFT®, designed by a vehicle during the winter), allow it to warm up Drive Medical GmbH &...
Page 3
b) Whilst raising or lowering Akkulift , do not place 1. Place hooked end of side flap protector over bath ® your hands between the edge of the bathtub and handle and pull downwards, as shown. Akkulift or under the seat of the chair, as this can ®...
Page 4
OPERATING THE Akkulift ® IN THE BATH 6. Take the hand grip with the other hand and fold up the backrest until it reaches the stop position of the 1. Before starting up your base-plate (see Figure 3). Akkulift , make sure that ®...
MANUAL CONTROL Emergency Stop-button (red): If there is any situation in which it is necessary to stop The green control lamp shows a fully charged battery. all functions of the bathlift, press the red emergency During the use the green control lamp of the manual button.
3. The backrest can now be removed from the frame in MAINTENANCE AND CARE the direction of the arrow shown in Figure 11. Note: The backrest cannot be removed from the frame MAINTENANCE OF THE AKKULIFT ® when it is completely folded down to the seat, because the backrest is then in a locked position again! is equipped with bearings and running sur- Akkulift...
Page 7
Taking care of your covers requires frequent cleaning, TRANSPORT either by hand or washing machine. They can be machine- Please observe the following instructions for mobile use washed up to 60 °C using a non-bio washing powder of the device: which does not contain any enzymes.
• Nicht aufladbare Batterien dürfen nicht aufgeladen werden. MAINTENANCE OF THE CHARGING UNIT • WARNING! Explosive gases may develop when char- The charging unit does not require any maintenance. ging lead batteries. Do not expose to open fire and The housing must only be cleaned with mild detergents, avoid sparks.
TROUBLESHOOTING ERROR POSSIBLE CAUSE REMEDY Lift does not work No storage battery in the battery com- Insert storage battery into the battery compartment partment Storage battery is empty Charge the storage battery Contacts are contaminated Remove the storage battery; clean the contacts with a dry cloth, and insert the storage battery again Manual control is not correctly connected Check the plug-type connection.
TECHNICAL DATA AKKULIFT ® Overall length (backrest in reclined position) 860 mm (32") Overall width (seat incl. side flaps) 700 mm (28") Width of seat 370 mm (14") Depth of seat 450 mm (18") Overall length of seat 580 mm (23") Sitting height min.
SPARE PARTS AND ACCESSORIES FOR AKKULIFT The correct order numbers of original spare parts and accessories are given in the table below. ARTICLE ORDER NUMBER SEAT AND BACKREST COVER Cover Set AKKULIFT® Riviera (UK) Classic, white 410 000 252 Cover Set AKKULIFT® Riviera (UK) Classic, blue 410 000 312 Cover Set AKKULIFT®...
Page 12
Please enter your specialist dealer´s address: ADDRESSES Drive Medical GmbH & Co. KG Drive DeVilbiss Healthcare Ltd Leutkircher Straße 44 Heathfield Lane, Birkenshaw, D-88316 Isny/Allgäu West Yorkshire, BD11 2HW Germany England Tel.: +49 (0) 7562 9724 - 0 Tel.: + 44 (0) 845 0600 333...
Need help?
Do you have a question about the Akkulift - Bathlift Riviera and is the answer not in the manual?
Questions and answers