Guidance And Manufacturer's Declaration; Electromagnetic Immunity - Datex-Ohmeda Aespire View User's Reference Manual

Table of Contents

Advertisement

Guidance and
manufacturer's
declaration -
electromagnetic
immunity
Power immunity
Immunity test
IEC 60601-1-2 test Level
Electrostatic
± 6 kV contact
discharge (ESD) IEC
± 8 kV air
61000-4-2
Electrical fast
± 2 kV for power supply
transient/burst IEC
lines
61000-4-4
± 1 kV for input/output
lines
Surge IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Voltage dips, short
<5% U
interruptions and
U
voltage variations on
40% U
power supply input
for 5 cycles
lines IEC 61000-4-11
70% U
for 25 cycles
<5% U
for 5 sec.
Power frequency (50/
3 A/m
60 Hz) magnetic field
IEC 61000-4-8 3
Note: U
is the AC mains voltage before application of the test level.
T
M1132382
The system is suitable for use in the specified electromagnetic
environment. The customer and/or the user of the system should
assure that it is used in an electromagnetic environment as described
below.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV differential mode
± 2 kV common mode
(>95% dip in
<5% U
T
) for 0.5 cycle
U
T
(60% dip in U
)
40% U
T
T
for 5 cycles
(30% dip in U
)
70% U
T
T
for 25 cycles
(>95% dip in U
)
<5% U
T
T
U
3 A/m
10 Specifications and Theory of Operation
Electromagnetic environment guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Mains power quality should be that of a
typical commercial and/or hospital
environment.
(>95% dip in
Mains power quality should be that of a
T
typical commercial and/or hospital
) for 0.5 cycle)
T
environment. If the user of the system
(60% dip in U
)
T
T
requires continued operation during
power mains interruptions, it is
(30% dip in U
)
T
T
recommended that the system be
powered from an uninterruptible power
(>95% dip in
T
supply or a battery.
) for 5 sec.
T
If display distortion or other abnormalities
occur, it may be necessary to position
the Anesthetic System further from
sources of power frequency magnetic
fields or to install magnetic shielding. The
power frequency magnetic field should
be immersed in the intended installation
location to assure that it is sufficiently
low.
10-27