Please contact a service technician or your retailer. Purposes and Applications HID-2138 is intended to be used in healthcare institutions for general purpose as an assisting device for data access – patient information, medical records, media services, and so on. The product is designed for general or special use in the hospital environment.
Page 3
3. Repair of the device may also only be carried out by Warning! Because of the danger of electric shock, never remove the cover of a device while it is in operation or connected to a power outlet. HID-2138 Quick Reference Guide...
Page 4
Position the power cord so that people cannot step on it. Do not place anything over the power cord. If the equipment is not used for a long time, disconnect it from the power source to avoid damage by transient over voltage. HID-2138 Quick Reference Guide...
Page 5
Medical electrical equipment Part 1-1: General requirements for safety Collateral standard: Safety requirements for Medical electrical systems – EN 60601-1-2 (IEC 60601-1-2) Medical electrical equipment Part 1-2: General requirements for safety Collateral standard: Electromagnetic compatibility; Requirements and tests HID-2138 Quick Reference Guide...
Page 6
Note! Environmental protection Follow national requirements to dispose of unit. Warning! Do not modify this equipment without authorization of the manufacturer. Warning! To avoid risk of electric shock, this equipment must only be connected to a supply main with protective earth. HID-2138 Quick Reference Guide...
Page 7
(if any) or even worse burn the equipment. 17. The product is not used in Category AP or Category APG in an anesthetic gas environment. HID-2138 Quick Reference Guide...
Page 8
Explanation of Graphical Symbols Warning: dangerous voltage Caution Note ISO 7000-1641: Follow operating instructions or Consult instructions for use. Direct current. Equipotential Stand-by US Conformance Follow the national requirements for disposal of equipment. Stacking layer limit This side up HID-2138 Quick Reference Guide...
Page 9
The mark on electrical and electronic products only applies to the current European Union Member States. HID-2138 Quick Reference Guide...
Page 10
(2) Use only shielded cables to connect I/O devices to this equipment. (3) Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. FCC RF Radiation Exposure Statement HID-2138 Quick Reference Guide...
Page 11
Follow national requirements to dispose of unit. Manufacturer Avalue Technology Inc. 7F, No.87, Lide St., Zhonghe District, New Taipei City, 235, Taiwan 7F, No.95, Lide St., Zhonghe District, New Taipei City, 235, Taiwan TEL: +886-2-8226-2345 FAX: +886-2-8226-2777 Web: www.avalue.com.tw Information: sales@avalue.com.tw HID-2138 Quick Reference Guide...
HID-2138 User Manual CONTENT 1. HID-2138 Multi Touch Medical Panel PC Features ........ 14 1.1 Packing List ..................14 1.2 Specifications ..................15 1.3 Front view ..................18 1.4 Rear & Bottom view ................19 1.5 System Dimensions ................20 2.
Page 13
5.3.25 Line In, MIC connector (JMIC1) ..........48 5.3.26 LPC connector (JLPC1) ............49 5.3.27 I2C EEPROM connector (JPD1) ........... 49 5.3.28 EDP connector (EDP1) ............50 6. General Safety Guide ................51 7. Guidance and Manufacturer’s Declaration ..........53 HID-2138 Quick Reference Guide...
HID-2138 User Manual 1. HID-2138 Multi Touch Medical Panel PC Features ▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬ In this chapter, you will get to know all features of our HID-2138 Multi Touch Medical Panel PC. 1.1 Packing List 1 x HID-2138 Medical Panel PC ...
DB-LDA001-4-060 Touch Screen 21.5 PCAP Touch Controller EETI EXC84H5680STAG External I/O Serial Port Optional COM port x 1—RS232/422/485 selected by BIOS USB 2.0 x 2 External USB Port USB type C x 2 (USB 3.1/Display/PD 3.0) HID-2138 Quick Reference Guide...
Page 16
2 Test frequency : 5~500 Hz 3 Sweep:1 Oct/ per one minute. (logarithmic) 4 Test axis : X,Y and Z axis 5 Test time :30 min. each axis 6 System condition : Non-Operating mode 7 Test curve HID-2138 Quick Reference Guide...
Page 17
-20°C ~ 60°C, 10% ~ 95% Condition (in range of °C and % R.H.) Atmospheric 700~1060hPa Pressure Expected Service 43368 Hrs Life Operating 40°C @ 95% Relative Humidity, Non-condensing Humidity Note: Specifications are subject to change without notice. HID-2138 Quick Reference Guide...
Volume Note1 Touch on/off* Up/down Up/down Note1: Press this button for 2 sec to turn on or off LED reading light bar Press this button for 4 sec to turn off or on for touch function HID-2138 Quick Reference Guide...
USB Type C: USB type C x 2 (Output for USB 3.1 & Display, Input for PD 3.0 15V) LAN: for internet connection Note! Equipotential terminal needs to be linked to the hospital ground/earth system before booting the system to protect both operator and system HID-2138 Quick Reference Guide...
HID-2138 User Manual 2. Setting Up HID-2138 Multi Touch Panel PCs ▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬ This chapter gives instructions on how to set up HID-2138 Multi Touch Panel PC and how to connect different cables. 2.1 VESA Mounting 2.2 Cleaning and Disinfecting HID-2138 Quick Reference Guide...
Installation instructions follow: First attach the wall-mounting to the heat-sink of the HID-2138, securing it in place with four of the M4 x 6mm screws provided. Mount the on the wall, stand or other flat surface.
Page 23
HID-2138 User Manual Suggested Screw type for mounting Note: 4 pieces of M4 x 6mm screws (P/N: E1933050611R) Warning! Use suitable mounting apparatus to avoid risk of injury. HID-2138 Quick Reference Guide...
Follow below step Connecting the Ground pin 1. With system ready, find the equipotential terminal on the rear side of the HID-2138. An equipotential terminal is provided to optionally connect to a hospital ground/earth system. 2. Prepare grounding cable and the other terminal links to the hospital ground/ earth system.
Page 25
HID-2138 User Manual Please follow below steps to connect power cable to system. The HID-2138 could only be powered by a DC power adapter (EDAC Model no. EM11013C). Be sure to always handle the power cords by holding the plug ends only.
HID-2138 User Manual 2.3 Cleaning and Disinfecting During normal use of HID-2138, the device may become dirty and should be regularly cleaned. Cleaning Instructions 1. Turn off the computer before starting clean up. This way, you can see any dirt on the screen; the brightness of the monitor may make you miss some areas.
HID-2138 User Manual 3. Using HID-2138 Multi Touch Panel PCs ▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬ This chapter describes in detail all features of HID-2138 Multi Touch Panel PC. 3.1 Turn ON/OFF the System 3.2 Using LCD Display and Touch Screen HID-2138 Quick Reference Guide...
1. Press on the Power ON/OFF icon firmly for 4 seconds. 2. The Power ON/OFF LED will turns orange to indicate power is off. 3. Your system is turned OFF. Note: We recommend using operating system shut down procedure to turn the system OFF. HID-2138 Quick Reference Guide...
3. Press LED reading light icon for 4 secs to turn off the touch function (touch function always on as default) 4. Press LED reading light icon for 4 secs to turn on the touch function HID-2138 Quick Reference Guide...
HID-2138 User Manual 4. HID-2138 Installation Figure ▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬ In this chapter, you will learn how to install storage into HID-2138. Please perform these steps with care. ▲ WARNING: Turn OFF the system and disconnect the power cable before performing the following tasks.
HID-2138 User Manual 4.1 WIFI dongle installation Step1. Remove 2 screws to release the chassis cover, and remove it. Step2. Insert the WIFI dongle. HID-2138 Quick Reference Guide...
HID-2138 User Manual 4.2 Desktop Stand installation Step1. Position VESA Mount on both sides, matching the holes on the system. Step2. Insert and fasten 4 screws on each side of the monitor to secure Mounting brackets. HID-2138 Quick Reference Guide...
HID-2138 User Manual 5.2 HID-2138 Jumper and Connector list Jumpers Label Function Note JBAT1 Clear CMOS 3 x 1 header, pitch 2.00mm JRI1/2 Serial port 1/2 pin9 signal select 3 x 2 header, pitch 2.00mm JSBKL1 LCD backlight brightness adjustment 3 x 1 header, pitch 2.00mm...
Page 35
2 x USB type C connector DCIN1 DC-IN connector JLPC1 LPC connector 5 x 2 header, pitch 2.00mm JPD1 I2C EEPROM connector 4 x 1 header, pitch 2.00mm M2B1 M.2 KEY-B1 connector M2B2 M.2 KEY-B2 connector USIM1 SIM card slot HID-2138 Quick Reference Guide...
HID-2138 User Manual 5.3.18 LED connector (JLED1) Signal +5VSB Reading_LED_EN 5.3.19 SATA Power connector (SPWR1) Signal +12V HID-2138 Quick Reference Guide...
5.3.20 Serial port 1 connector (JCOM1) Signal PIN PIN Signal COM_DCD#_1 COM_RXD_1 COM_TXD_1 COM_DTR#_1 COM_DSR#_1 COM_RTS#_1 COM_CTS#_1 COM_RI#_1 5.3.21 Serial port 2 connector (JCOM2) Signal PIN PIN Signal COM_RI#_2 COM_CTS#_2 COM_RTS#_2 COM_DSR#_2 COM_DTR#_2 COM_TXD_2 COM_RXD_2 COM_DCD#_2 HID-2138 Quick Reference Guide...
HID-2138 User Manual 5.3.22 Serial port 1 in RS-422/485 mode (JRS485_1) Signal PIN PIN Signal 422RX1+ 485_422TX1+ 422RX1- 485_422TX1- 5.3.23 Serial port 2 in RS-422/485 mode (JRS485_2) Signal PIN PIN Signal 485_422TX2- 422RX2- 485_422TX2+ 422RX2+ HID-2138 Quick Reference Guide...
HID-2138 User Manual 5.3.24 Front Audio connector (JFAUD1) Signal PIN PIN Signal HP_L HP1_JD MIC1_JD HP_R ACZ_DET# MIC1_RIN MIC1_LIN 5.3.25 Line In, MIC connector (JMIC1) Signal MIC2_RIN HID-2138 Quick Reference Guide...
You want to remove/install any parts Thermal The HID-2138 is a fanless design system, heat is dispatch through rear metal heatsink which is located at VESA mount area.. When using your HID-2138 systems, it is normal for the metal heatsink to get warm. The rear metal heatsink of the HID-2138 functions as a cooling surface that transfers heat from inside the computer to the cooler air outside.
Page 52
Handle your HID-2138 with care. It is made of metal, glass, and plastic and has sensitive electronic components inside. Don't use a damaged HID-2138, such as one with a cracked screen, as it may cause injury. Setup HID-2138 on a stable work surface.
Portable and mobile RF communications equipment can affects system. See tables 1, 2, 3, and 4 below. HID-2138 Quick Reference Guide...
Page 54
Compliance Emissions power supply network that supplies IEC 61000-3-3 buildings used for domestic purposes HID-2138 Quick Reference Guide...
Page 55
70% U (30% in U ) for 25 cycle or a battery. dip in U ) for 25 cycle <5% U (>95% dip in U ) for 5 <5% U (>95% dip in U ) for 5 HID-2138 Quick Reference Guide...
Page 56
A.C. mains voltage prior to application of the test level. HID-2138 Quick Reference Guide...
Page 57
Field strengths from fixed RF IEC 61000-4-3 2.5GHz transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: HID-2138 Quick Reference Guide...
Page 58
Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 1V/m. HID-2138 Quick Reference Guide...
Page 59
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. HID-2138 Quick Reference Guide...