Emergency Reserve Manifold
CAUTION: Supplied fixings are for use
with solid masonry type walls only. Typical ERM
weight is 16kg.
Note - 'J' type cylinders typically for Oxygen and Medi-
cal Air.'G' type cylinders typically for Nitrous Oxide
and N2O/O2 mix (Entonox).
2.1.8
The pipework should be secured to the
wall using munsen rings (not supplied). It would
be recommended to fit the first pipe support to
the supplied ø22mm OD stub pipe (Item 12, figure
1). The next support should typically be fitted
within 2m of the first.
The ø28mm exhaust line (Item 13, figure
2.1.9
1) shall be brazed using fluxless brazing technique
with nitrogen purge.
CAUTION: The ø28mm exhaust line (Item
13, figure 1) needs to be piped away from the
manifold room to a safe location to prevent
buildup of waste gas in an enclosed space in the
event of a regulator failure.
CAUTION: Do not reduce the diameter of
the pipe used for the exhaust line.
Figure 4 - Typical Installation For use with 'VF'
Type Cylinders
CAUTION: Supplied fixings are for use
with solid masonry type walls only. Typical ERM
weight is 16kg.
Note - 'VF' size cylinders typically used for CO2
9
2005715.05