Emergency Reserve Manifold
0. Safety Precautions
This section gives safety, storage and handling
information for the BeaconMedæs Emergency
Reserve Manifold only. Component parts lists and
descriptions are available on request.
Operators should have carefully read and become
familiar with the contents of this manual before
maintaining the Emergency Reserve Manifold.
Operators are expected to use common sense
safety precautions, good workmanship practices
and follow any related local safety regulations.
0.1 Identification of symbols
The following symbols apply to this product and
are used in these instructions and on the product
in question. The meanings of these symbols are as
specified below: -
Read instructions
Ambient temperature range
Ambient humidity range
Ambient pressure range
Date of manufacture
Do not dispose of in general waste
0.2 Environmental Transport and Storage
Conditions
All products are separately packaged and stored in
controlled conditions.
0.3 Environmental Operating Conditions
Adverse environmental conditions and harsh
abrasives or chemicals may cause damage to the
unit.
0.4 Environmental Protection
Discard the unit and/or components in any
standard refuse facility. The unit does not contain
and hazardous substances.
0.5 Electromagnetic Interference
2005715.05
Not applicable
0.6 Cleaning
The manifold should be wiped over with a damp
cloth frequently to remove any dust or foreign
substances
0.7 Safety Notice
Persons undertaking any installation and/or
maintenance must be fully trained in specialist
work of this nature.
The "PERMIT TO WORK" procedure must be
adhered to for all installations once commissioned.
The manifold is designed and built in accordance
with HTM 02-01 regulations and therefore should
be installed as such.
Oil, grease and jointing compounds must not be
used.
Do not attempt to prove the pressure relief valve,
under any circumstances, by altering the regulator.
Pressure relief valves must be removed and tested
off site by a registered test centre and a certificate
of conformity issued.