11. Guidance and Manufacturer's Declaration
Electromagnetic Compatibility
Important information regarding Electromagnetic Compatibility
(EMC)
M7 Intelli IT (ALRU) conforms to EN60601-1-2:2015
Electromagnetic Compatibility (EMC) standard.
Nevertheless, special precautions need to be observed:
•The use of accessories and cables other than those specified or
provided by OMRON could result in increased electromagnetic
emission or decreased electromagnetic immunity of the device and
result in improper operation.
•During measurement, the use of the device adjacent to or stacked
with other device should be avoided because it could result in
improper operation. In case such use is necessary, the device and
other device should be observed to verify that they are operating
normally.
•During measurement, Portable RF communications device
(including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm to any part of the
device, including cables specified by OMRON. Otherwise,
degradation of the performance of the device could result.
Table 1 - EMISSION Limits and Compliance
Phenomenon
Conducted and
radiated RF
EMISSIONS
Voltage fluctuations
and flicker
EN61
EMISSION Limits
CISPR 11
See IEC 61000-3-3
Conformity
Group1, Class B
Complies