Movement error symbol
Appears with your reading when your body is moving
during a measurement. The body movement
function is disabled when the possibility of Afib or
R
irregular heartbeat is detected during a Afib mode
measurement. If it appears, remove the arm cuff and
wait for 2 to 3 minutes. Then, reapply the cuff,
remain still and take another measurement.
Irregular heartbeat symbol
Appears when irregular rhythm is detected 2 or more
times during a measurement.
It does not appear during a Afib mode measurement.
An irregular heartbeat rhythm is defined as a rhythm
that is 25 % less or 25 % more than the average
rhythm detected while your monitor is measuring
blood pressure.
S
Irregular heartbeat symbol may be displayed when
the monitor detects arm movements. Refer to
section 7 for more information.
If it continues to appear, we recommend you to
consult with and follow the directions of your
physician.
Cuff wrap guide symbol (loose)
Appears if the arm cuff is not wrapped around the
T
arm correctly while taking a measurement. It also
appears when viewing past readings.
Pulse display / Interval counter for Afib
Pulse rate appears after the measurement. Counts the seconds
U
for next Afib measurement during Afib mode.
Irregular Heartbeat
Normal Heartbeat
1. Know Your Monitor
Long
Short
EN