Single And Multiple Buccal Registration - Dentsply Sirona Primescan Connect Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Primescan Connect
67 97 216 D3801
D3801.201.01.03.02    2022-09
5.2.5

Single and multiple buccal registration

The buccal registration establishes the allocation of jaw exposures.
ü The jaw with the preparation is scanned.
1. Scan the occlusal, buccal and lingual view of the antagonist (see
the section "Occlusal scan  [→ 39]", "Buccal scan  [→ 39]" and
"Lingual scan  [→ 40]").
2. Perform a buccal scan of the bite block prior to completing the
registration. This buccal scan should be carried out close-up to the
preparation. To acquire sufficient geometry, capture the teeth of the
upper and lower jaw as well as 5 mm of the respective gingival
areas.
3. Please complete a buccal scan on both sides for a full jaw scan. For
this, use the scanner to complete a buccal scan in each case over
the premolars of both quadrants.
Tip: In the case of multiple or long-span restorations over several
quadrants, we recommend generating several buccal exposures close
to the restoration.
5 Operation
5.2 Scanner guide
41