Occlusal Scan; Buccal Scan - Dentsply Sirona Primescan Connect Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Primescan Connect
67 97 216 D3801
D3801.201.01.03.02    2022-09
5.2.1

Occlusal scan

Important: Ensure that the distance between the coated sapphire glass
of the scanner and the scanned surface is observed.
The distance must be between 0  -   2 0 mm (ideally: 2  m m). The scanner
does not rest on the teeth or the gums.
1. Move the scanner to the starting position. For this purpose, the
scanner is in the occlusal view of the tooth, which is next to the
prepared tooth in the distal direction.
2. Scan in the mesial direction. To do so, move the scanner in the
occlusal direction from the distal-positioned tooth over the prepared
tooth to the mesial-positioned tooth.
5.2.2

Buccal scan

ü The scanner is on the adjacent tooth, in the mesial direction to the
preparation.
1. Rotate the scanner 20° toward the buccal.
2. Guide the scanner over the entire buccal distance in the distal
direction over the prepared tooth.
5 Operation
5.2 Scanner guide
39

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents