Table 7-5: Error Codes 400-499: Motion Errors
Error
Code
400
401
402
403
Table 7-6: Error Codes 500-599: NVRAM Errors
Error Code
500
501
502
503
504
505

Opening the Drawer Manually

Do the following if an error occurs while the drawer is closed and you need to remove a plate.
Press Drawer.
If the drawer does not open after you press Drawer, then turn off the power to the
instrument and then on again.
If the drawer remains closed after you cycle the power, and the incubator is on, then
turn off the incubator.
If the drawer still remains closed, power off the instrument, unplug the power cord and
then at the groove in the upper left side wall of the drawer, try to use a blunt, flat object,
such as a spatula, to open the door. With your index finger, pull the plate drawer out of
the instrument (do not force the drawer) and remove the plate. This action will not harm
the instrument, but should only be taken if the first two options fail to open the drawer.
If you are still unable to open the drawer, contact your local Molecular Devices
representative.
0112-0102 E
Error Message Notes
carriage
Carriage did not move to either of its photo interrupts in a reasonable
motion error
time, or cannot find its photo interrupt.
filter wheel
Filter wheel did not move to its photo interrupt in a reasonable time, or
error
cannot find its photo interrupt.
grating error
Grating did not move to its photo interrupt in a reasonable time, or
cannot find its photo interrupt.
stage error
Stage did not move to its photo interrupt in a reasonable time, or cannot
find its photo interrupt.
Error Message
NVRAM CRC corrupt
NVRAM Grating cal data bad
NVRAM Cuvette air cal data error
NVRAM Plate air cal data error
NVRAM Carriage offset error
NVRAM Stage offset error
Chapter 7: Troubleshooting
Notes
The CRC for the NVRAM data is corrupt.
Grating calibration data is unreasonable.
Cuvette air calibration data is unreasonable.
Plate air calibration data is unreasonable.
Carriage offset data is unreasonable.
Stage offset data is unreasonable.
43