SpectraMax M2 and M2e Microplate Reader User Guide
Table 7-2: Error Codes 100-199: Unrecognized Command Errors Sent From the Computer
Error
Code
100
101
102
103
104
105
106
107
108
109
110
111
40
Note: If an error occurs while the plate drawer is closed, and you need to remove a
plate, see
Opening the Drawer Manually on page
Error Message
Notes
command not
Command string not recognized.
found
invalid argument
Command argument not recognized.
too many
Too many arguments after command.
arguments
not enough
Missing arguments.
arguments
input line too long
Too many characters in the input line.
command invalid,
Instrument could not perform the give command because it was busy
system busy
doing another task.
command invalid,
Instrument could not perform command because a measurement was
measurement in
in progress.
progress
no data to transfer Inputting transfer when there is no data in the buffer.
data buffer full
Too many data sets in the buffer. Can be caused by setting up a long
kinetic read and then disconnecting computer or the SoftMax Pro
Software is preempted by another application.
error buffer
More than 65 errors in the buffer, clear the buffer.
overflow
stray light cuvette,
Cuvette door open while doing a read.
door open?
invalid read
settings
43.
0112-0102 E
Need help?
Do you have a question about the SpectraMax M2 and is the answer not in the manual?