8. Run a single stream analysis on the stream associated with the HCG until the readings stabilize.
9. Run a forced calibration on the stream associated with the HCG.
10. Run a normal calibration on the stream associated with the HCG.
The GC is ready to analyze the sample or validation stream using the 2 Pt Exp with the response factor that was calculated
during the LCG and HCG runs.