Download Print this page

Emerson Rosemount 700XA Reference Manual page 22

Gas chromatograph
Hide thumbs Also See for Rosemount 700XA:

Advertisement

July 2020
Front panel assembly
The front panel assembly is located on the front section of the lower enclosure and consists of a removable, explosion-proof panel
that shields either a switch panel or a local operator interface (LOI).
Figure 13: 8-stream switch panel (left) and 18-stream switch panel (right)
Switch panel
The switch panel contains a network of on/off switches that allow you to manually control the GC's stream and analytical valves.
Figure 14: 8-stream Switch Panel
There are two types of switch panels: 8-stream and 18-stream. The 8-stream switch panel is the standard panel, and is used when
the GC has only one heater/solenoid board installed; if two heater/solenoid boards are installed, then the 18-stream switch panel is
used.