Care And Maintenance; Periodic Maintenance; Store - sauter HMO Instruction Manual

Mobile leeb hardness tester
Hide thumbs Also See for HMO:
Table of Contents

Advertisement

5 Care and maintenance

5.1 Periodic maintenance

General information
Avoid shocks. After use, place the device back into the transport case.
The rebound sensor must be stored in a triggered state.
Avoid use in a strong magnetic field. Protect all components from any contact with
grease or oil.
Clean impact device
In principle, the tube and the housing of the impact device must be cleaned every 1 to
2 thousand measurements.
Replacement of the impact ball
Continuous use may damage the impact ball. It shall be replaced as soon as its
condition appears to affect the accuracy of the measurement.