Keysight Technologies N7700 User Manual page 36

Photonic application suite
Table of Contents

Advertisement

3
Fast Spectral Loss Measurement Application
36
The average power should serve as a verification for the reference
NOTE
validity. Depending upon the TLS power selected in the GUI and the
insertion loss of the reference connection, you should compare this value
with your expectation of the average power.
You can perform a single reference sweep on multiple ports
simultaneously by using the check boxes above the individual ports. All
checked ports will be included in the reference sweep, once you click the
New Ref. (sel.) button.
To obtain a reference on all ports simultaneously, click the New ref. (all)
button. In this case, it doesn't matter which ports have been checked or
unchecked.
Use the Select All and Select None buttons to set or remove all check
marks simultaneously.
You can obtain reference data on one or more ports, then copy those
traces to be used as reference for other ports as well or instead. To do so,
obtain the actual reference sweep, then check all destination ports (make
sure all other ports are unchecked) and click the Cpy Frm button at the
source port.
All destination ports will show the detail information of the source port
afterwards.
You can clear data for individual ports by clicking the Clear button of the
respective port. By checking several ports and clicking the Clear selected
button, reference data for those ports is cleared. By clicking the Clear Ref.
button, the complete reference data is cleared from memory. Also upon
restarting the engine, no reference will be loaded automatically, until a
new reference has been saved.
Ports that don't display reference details (such as timestamp, etc.) will
NOTE
be referenced to the TLS power set in the Measurement Parameters
section.
You can load, save and delete reference files by clicking the Load Ref.,
Save Ref. and Delete Ref. buttons. By clicking either button, a list of
reference files will be displayed to choose from.