8.3

Autoclavable Components

Autoclavable Components
Heating needle
WARNING
Only the components above can be autoclaved.
Before first use and after each use, sterilize the above components.
Reprocessing Instructions
Preparation
at the Point
of Use:
Transportat
ion:
Preparation
for
Decontamin
ation:
Pre-
Cleaning:
8 Maintenance
Before cleaning, disconnect the heating needles from the
handpiece. Refer to Chapter 4.1 of this manual for disassembly
instructions. Remove gross contaminations from the components
with code water (<40°C) immediately after use. Don't use a fixating
detergent or hot water (>40°C) as this can cause the fixation of
residuals which may influence the result of the reprocessing
process.
Store the instruments in a humid surrounding.
Do not submerge the components or wipe them with any of the
following functional water (acidic electrolyzed water, strong
alkaline solution, or ozone water), medical agents (glutaral, etc.),
or any other special types of water or commercial cleaning liquids.
Such liquids may result in metal corrosion and adhesion of the
residual medical agents to the components.
Safe storage and transportation to the reprocessing area to avoid
any damage and contamination to the environment.
The devices must be reprocessed in a disassembled state.
Observe suitable personal protective measures.
Do a manual pre-cleaning, until the components are visually clean.
Submerge the components in a cleaning solution and flush the
lumens with a water jet pistol with cold tap water for at least 10
18 / 31