Camera Rotation - Siemens SIREMOBIL Compact L Service Instructions Manual

Table of Contents

Advertisement

5 - 20
Fig. 3
Fig. 4

Camera rotation

WARNING
NOTE
NOTE
SIREMOBIL Compact
System Manual
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
The keys for horizontal and vertical image reversal must be off
prior to adjusting camera rotation. Camera rotation should be set
in the 0° position. Fig. 3, Fig. 4 and Fig. 5 show the 0° position.
With the right camera rotation key (the arrow on the key is point-
ing clockwise), the marker on the monitor will rotate clockwise.
The image contents will rotate counterclockwise.
With the left camera rotation key (the arrow on the key is pointing
counterclockwise), the marker superimposed on the monitor will
rotate counterclockwise. The image contents will rotate clock-
wise.
SPR2-130.061.02
Rev. 11
Adjustments / Programs
Patient left position
Patient right position
Fig. 5
Page 20 of 30
09.04
CS PS 24
5
Siemens AG
Medical Solutions

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents