12. Technical Specifications

Operating conditions: 10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
-20 - +55 °C / -4 - +131 °F
Storage conditions:
15 - 95 % relative maximum humidity
340 g (including batteries)
Weight:
136 x 82 x 57 mm
Dimensions:
Measuring procedure: oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
0 - 299 mmHg
range:
1 mmHg
Resolution:
pressure within ± 3 mmHg
Static accuracy:
± 5 % of the readout value
Pulse accuracy:
4 x 1.5 V alkaline batteries; size AA
Voltage source:
Mains adapter DC 6V, 600 mA (optional)
approx. 920 measurements
Battery lifetime:
(using new batteries)
IP20
IP Class:
EN 1060-1 /-3 /-4; IEC 60601-1;
Reference to
IEC 60601-1-2 (EMC); IEC 60601-1-11
standards:
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.