SAFETY INFORMATION

Fig. 1
Applicable bulb replacement: 6V30W Halogen bulb P/N EL-455
3. Install microscope on a sturdy, level table or bench and avoid any restriction of air vents in the base of
the unit.
Do not place microscope on a flexible surface, as this could result in blocking the air vents and cause
overheating.
4. Always use the power cord provided by LABOMED. If the proper power cord is not used, product safety
performance cannot be warranted.
5. When installing the microscope, route the power cord away from the microscope frame. Should the
power cord come in contact with the microscopes base, the power cord could melt due to overexposure
to heat.
6. Always ensure that the grounding terminal of the microscope and that of the wall outlet are properly
connected. If the unit is not grounded, LABOMED can not warrant electrical safety.
7. Never allow metallic objects to penetrate the air vents of the microscope frame as this could result in
user injury and damage to the microscope.
8. After operation of microscope, be sure to disconnect power cord from connector socket of the microscope
or from the wall power outlet.
SAFETY SYMBOLS
The following code categories describe the degree of danger likely to be incurred in the event of user error
made in ignorance of these symbols.
Symbol
Hot parts symbol.
This symbol is placed on bulb housing and on other parts to indicate that they may be hot. Never
touch those parts just after the equipment has been used. Allow sufficient time to cool before
touching them. In the event of user error it is possible to injury.
In the event of user error, death or serious injury is possible.
WARNING
!
In the event of user error, product may be damaged.
ATTENTION
The following describes icon directives
!
This icon denotes "Refer to User Manual" and "Caution".
This icon represents the productive earth.
This icon represents alternating current.
7125000-795
1. After the microscope has been used for observation of a speci
-men containing bacteria, clean all parts coming in contact with
the specimen to prevent infection.
• Be sure to remove the specimen before moving this product.
• In case the specimen is damaged by erroneous operation, it is
important to clean all surfaces that may have come in contact
with the specimen.
2. To avoid potential electrical hazards when replacing either halo
-gen bulb, turn the microscope main switch to the OFF posi
-tion and disconnect power cord from wall outlet in advance.
Whenever you replace your microscope bulb, allow lamp socket
and bulb to cool before touching (Fig.1)
Explanation
TCM 400
TCM 400
2
Issue 1.3
Printed on April, 2022