Description of the Interface
-
-
-
Features
-
Type of interface: Serial interface
Operating mode: Full duplex
-
Standard: RS-232
-
Transmission rates: 600, 1200, 2400, 4800, 9600, 19200,
-
38400, 57600 and 115200 baud
Parity: odd, even, none
-
Number of data bits: 7 or 8 bits
Character format:
1 start bit, 7-bit ASCII, parity, 1 Stop bit
-
Handshake:
For 2-wire interface: Software
For 4-wire interface: Hardware (CTS/RTS)
Data output format of the balance:
16 or 22 characters
-
-
-
Factory Settings
-
: 9600 baud (Code 2.1.2.7.)
Transmission rate
Parity: ODD (Code 2.1.3.3.)
-
Stop bits: 1 STOP BIT (Code 2.1.4.1.)
-
Handshake:
HRDWARE. Hardware (Code 2.1.5.2.)
Operating mode: SBI (Code 2.1.1.1.)
Printing: IND.NO. Manual without stability (Code 3.1.1.1.)
Preparation
See "Pin Assignments" and "Pin Assignment Chart"
Technical Note
Your balance is equipped with a
serial RS232 interface port for
connection to a computer or
other peripheral device.
You can connect a computer to
change, start and/or monitor the
functions of the balance and the
application programs.

Advertisement

loading

Summary of Contents for Sartorius Entris II

  • Page 1 Entris ® Description of the Interface Technical Note Your balance is equipped with a serial RS232 interface port for connection to a computer or other peripheral device. You can connect a computer to change, start and/or monitor the functions of the balance and the application programs.
  • Page 2: Configuring The Interface

    Configuring the Interface Parameter Settings (Menu): Please refer to the installation and operating instructions supplied with your balance. Data Output Format with 16 Characters Display segments that are not activated are output as spaces. The type of character that can be output depends on the character’s position: Position –...
  • Page 3 Data Output Format with 22 Characters When data is output with an ID code, the 6-character code precedes the 16-character string described above. The code identifies the subsequent value. – ID code character Unit symbol ¹ Space CR: Carriage Return D: Digit or letter LF: Line Feed Example:...
  • Page 4 Commands (Data Input Format) You can connect a computer to your balance to send commands via the balance interface port for con- trolling balance functions and applications. The commands sent are control commands and may have different formats. Control commands consist of up to 13 characters.
  • Page 5: Pin Assignment Chart

    The conditions specified by Sartorius. Failure to do so may that can be configured are listed in the descriptions of the damage or even completely ruin your balance and/or periph- application programs.
  • Page 6: Cabling Diagram

    Cabling Diagram For connecting a computer or other peripheral device to the balance using the standard DCE protocol and cable lengths of up to 15 m (approx. 50 ft). Important: do not connect any other pins to the cable connector of the balance. Waage Computer Balance...
  • Page 8 Bohemia, NY 11716 Phone +49 551 308 0 Phone +1 631 254 4249 Toll-free +1 800 635 2906 For further contacts, visit www.sartorius.com Specifications subject to change without notice. Copyright Sartorius Lab Instruments GmbH & Co. KG. Status: 01 | 2021...

This manual is also suitable for:

Bce653-1nusBce323-1nusBce623-1nusBce3202-1nusBce4202-1nusBce8201-1nus ... Show all