Martin maxium Operating Instructions Manual page 89

Table of Contents

Advertisement

Martin maxium & maxium Beamer Operating Instructions
Protection class
Classified acc. to MDD
Leakage currents (LF/HF)
Type (of applied part)
Duty type
Device fuses
Sound level
Weight
Interference suppression
Dimensions
Environmental conditions
for transportation and stor-
age
Environmental conditions
for operation
V 2.0
I acc. to IEC 601-1
II b
in compliance with IEC 601,
CF; defibrillation-proof
intermittent 10 s / 30 s = duty factor of 25%
2 x T 6.3 A (slow-blow)
HF current activation: 55 dB(A) (adjustable 45–60 dB)
Alarm: 65 dB(A)
8.3 kg
Limits in conformity with EN 55011 and IEC 601-2-2
Interference immunity in conformity with IEC 1000
Width
390 mm
Height
182 mm
Depth
435 mm
Ambient temperature
Relative humidity
Atmospheric pressure
Ambient temperature
Relative humidity
Atmospheric pressure
in conformity with Directive 93/42/EEC
-25°C to +70°C
10% to 100%
500 hPa to 1060 hPa
+10°C to +40°C
30% to 75%
700 hPa to 1060 hPa