Martin maxium Operating Instructions Manual page 88

Table of Contents

Advertisement

Current Type
Pulsed Argon: slow repetition rate*
Pulsed Argon: fast repetition rate *
Bipo Pure Cut
Bipo Blend Cut
Forfex
Micro Coag
Micro Stop
Auto Micro
Macro Coag
Macro Stop
Auto Macro
Seal Safe
Endo Seal
V
V
Note:
= peak voltage;
p
* Data applies to coagulation phase only; cycle-based average value deviates.
** Data averaged across cycle; pulse-based average value deviates.
n.a. = not applicable
Only the following currents are permitted for use with the baby NE (maximum power indicated in each case):
Pure Cut 100 W
Blend Cut 100 W
Super Blend 100 W
Argon Cut 100 W
88
HF Power
1–100 W at 500 ohms (CF 18–5.4) max. 4,000 V
1–100 W at 500 ohms (CF 18–5.4) max. 4,000 V
1–120 W at 300 ohms
1–120 W at 300 ohms
1–120 W at 50 ohms
0.1–50 W at 200 ohms
0.1–50 W at 200 ohms
0.1–50 W at 200 ohms
1–120 W at 50 ohms
1–120 W at 50 ohms
1–120 W at 50 ohms
5–320 W at 25 ohms
1–100 W at 50 ohms
= peak to peak
pp
Contact Coag 90 W
Contact Coag Auto Stop
90 W
Spray 80 W
Endo Spray 100 W
Martin maxium & maxium Beamer Operating Instructions
Crest Factor
Argon Beam 100 W
Argon Endo 100 W
Care Coag 50 W
Care Coag Auto Stop 50 W
Care Cut 70 W
HF Voltage
CF 1.7
max. 600 V
CF 2.1
max. 850 V
CF 1.9
max. 180 V
CF 1.8
max. 200 V
CF 1.8
max. 200 V
CF 1.8
max. 200 V
CF 1.9
max. 180 V
CF 1.9
max. 180 V
CF 1.9
max. 180 V
CF 1.9
max. 300 V
CF 2.1
max. 230 V
Clamp Coag 90 W
Forced Clamp 90 W
Pulsed Argon: slow
repetition rate 100 W
Pulsed Argon: fast
repetition rate 100 W
max. 5,900 V
p
pp
max. 5,900 V
p
pp
max. 1,200 V
p
pp
max. 1,700 V
p
pp
max. 360 V
p
pp
max. 400 V
p
pp
max. 400 V
p
pp
max. 400 V
p
pp
max. 360 V
p
pp
max. 360 V
p
pp
max. 360 V
p
pp
max. 600 V
p
pp
max. 450 V
p
pp
V 2.0

Advertisement

Table of Contents
loading

This manual is also suitable for:

Maxium beamer

Table of Contents