Important information About this document Other symbols These symbols are used in the document and These installation and operating instructions on or in the unit: form part of the unit� Note, e�g� specific instructions regarding If the instructions and information in efficient and cost-effective use of the unit�...
Important information Safety 1.2 Copyright information All names of circuits, processes, names, soft- Dürr Dental has designed and constructed this ware programs and units used in this document device so that when used properly and for the are protected by copyright� intended purpose there is no danger to people The Installation and Operating Instructions must or property�...
Important information 2.5 Qualified personnel 2.8 Transport The original packaging provides optimum pro- Operation tection for the device during transport� Persons who operate the units must ensure safe If required, original packaging for the unit can be and correct handling based on their training and ordered from Dürr Dental�...
Product description 3.1 Scope of delivery 3.3 Disposable materials The following items are included in the scope of The following materials are consumed during delivery (possible variations due to country-spe- operation of the device and must be ordered cific requirements and/or import regulations): separately: Orotol plus Variosuc VSA .
Product description Technical data Electrical data for the device Nominal voltage V AC Voltage fluctuation ±10 Nominal current 2�9 Frequency Type of protection IP20 Protection class Duty cycle 100 (S1) G-fuse link IEC 60127-2 T6�3 AH Unit plug 1/N/PE General technical data Dimensions (W x H x D) 360 x 1000 x 625 Weight...
Product description Operation 4.1 Type plate The type plate is located on the rear of the unit� 4.2 Conformity assessment This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines� This equipment conforms to all relevant require- ments�...
Installation Requirements Installation 6.1 Installation/setup room 7.1 Setting up the unit The room chosen for set up should fulfil the Where a waste water connection is available, following requirements: lead the fluids directly via the waste water hose into the waste water outlet� –...
Installation 7.3 Electrical safety when making Remove the protective cover� connections The unit must only be connected to a correct- ly installed power outlet� Do not place non-fixed multi-socket units on the floor� Follow the requirements in sec- tion 16 of IEC 60601-1 (EN 60601-1)� Do not operate any other systems using the same multiple socket�...
Installation 7.4 Connecting the unit to the mains Requirements: – Properly installed power outlet close to the unit (max� mains cable length 2�5 m)� – Easily accessible power outlet� – Mains voltage must match the information shown on the type plate� Further information is provided in the installation and operating instructions of the combination suction units VS 300 S and VSA 300 S�...
Installation Commissioning and first start-up NOTICE Short circuit due to the build up of condensation Do not switch on the unit until it has warmed up to room temperature and it is dry� Two labels are included in the scope of delivery of the bacteria filter�...
Operation Operation 10 Display panel 9.1 Aspirate fluid Remove the hose from the hose manifold to aspirate� The combined suction unit is started� Aspirate fluid from the patient's mouth� The fluid is collected in the fluid container� After treatment, check the filling level of the fluid container�...
Operation 10.3 Amalgam collector vessel is 10.5 Motor fault 100% full Red display and Yellow LED is on green LED flash alternately Red display flashes Audible signal Audible signal melody sounds Occurs during the start-up of the amal- gam separator� –...
Operation 11 Cleaning and disinfection 11.2 After every treatment Aspirate a glass of cold water through the The reprocessing of the handpieces and large and the small suction hoses� Do this the cleaning of the hose manifold is de- even if only the small suction hose was actual- scribed in the installation and operating ly used during treatment�...
Operation 11.5 Weekly and before longer treat- Place the rinsing hose onto the connector on the cover of the fluid container and secure ment interruptions with the clip� Flush the unit at least weekly and before longer treatment interruptions� Wear protective equipment to avoid any risk of infection (e�g�...
Page 19
Operation Place the large suction handpiece onto the rinsing hose� After c� 20 min� remove the suction handpiece slowlyfrom the connection of the rinsing hose,and place in the hose manifold� Remove the rinsing hose from the fluid con- tainer and clean and disinfect using a suitable instrument disinfectant, e�g�...
Operation 12 Maintenance Only trained specialists or personnel trained by Dürr Dental may service the unit� Prior to working on the device or in case of danger, disconnect it from the mains (e� g� pull the mains plug)� Maintenance Maintenance work interval Weekly Change disposable filter�...
Operation 12.1 Grease the seal 12.2 Replace the disposable filter Grease the seal if required� NOTICE Risk of equipment damage Do not operate the unit without the dis- posable filter� Wear hand protection� Open the cover and remove the disposable fil- ter�...
Page 22
Operation Disconnect the bacteria filter from the con- necting sleeve� 9000-606-40/30 1707V001...
Appendix 13 Information about EMC in accordance with EN 60601-1-2 13.1 General information The information in this leaflet includes excerpts from the relevant European standards for electrical, medical devices� It must be observed when installing Dürr Dental devices or combining them with products of other manufacturers�...
Page 24
Appendix Resistance to electromagnetic interference (immunity) for all devices and systems The device is designed for use in electromagnetic environments specified below� The customer or op- erator of the device should ensure that the device is operated such an environment� Interference immu- IEC 60601 - test Compliance level...
Page 25
Appendix Electromagnetic interference immunity for devices or systems that are not life-sustaining Portable and mobile communication devices should not be used any closer to the unit (including ca- bles) than the recommended safety distance, which is calculated based on the applicable formula for the transmission frequency�...
Page 26
Appendix Recommended safety distance between portable and mobile HF communication devices and the unit The device is designed for use in the electromagnetic environments specified below, in which the HF disturbance variables are controlled� The customer or the operator of the device can help to prevent electromagnetic interference by maintaining the minimum distances between mobile HF communica- tion equipment (transmitters) and the device as recommended below in accordance with the maxi- mum output line of the communication equipment�...
Page 28
Hersteller/Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...