Cleaning; Before Cleaning; External Surfaces; Internal Components - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Cleaning

Before Cleaning

External Surfaces

Internal Components

Revision C
Maintenance: Cleaning
:$51,1*
SHOCK HAZARD — In order to remove power from the
ApexPro Telemetry Server, disconnect the power cord from the
wall outlet. The on/off button on the front panel does not
disconnect the ApexPro Telemetry Server from AC power.
Additionally, wait after disconnecting the ApexPro Telemetry
Server power cord from the wall outlet. The ApexPro Telemetry
Server remains energized for a period of time after shutdown.
Failure to follow these instructions could result in electrical shock.
1.
Shutdown the ApexPro Telemetry Server according to
ApexPro Telemetry Server"
2.
Turn off and unplug the ApexPro Telemetry Server.
127(
When unplugging, pull on the plug, not the cord.
&$87,21
EQUPMENT DAMAGE/SHOCK HAZARD —Make sure water
does not run into any of the openings of the ApexPro Telemetry
Server. An electrical shock hazard exists and/or damage may
result if water is allowed to get inside the ApexPro Telemetry
Server. Failure to follow these instructions could result in
equipment damage and/or electrical shock.
Use a soft, lint-free cloth dampened with clean water to wipe down the external
surfaces of the ApexPro Telemetry Server. Do NOT use aerosol liquid sprays,
solvents, or abrasives.
1.
Verify the ApexPro Telemetry Server is turned off and unplugged.
2.
Disconnect all cords/cables from the ApexPro Telemetry Server.
3.
Remove the access panels on the ApexPro Telemetry Server. Using a
compressed air duster recommended for computers and electronics, carefully
blow out any accumulated dust.
ApexPro Telemetry Server
2001989-178
"Stop/Start/Restart
on page 2-29.
4-7

Advertisement