Receiver Function/Available Receivers Check

Revision C
Maintenance: Checkout
To run disk check:
1.
Select Start > Shut Down > Log off, then click OK.
2.
Log back in as administrator, according to
Logons"
on page 2-18.
3.
After the ApexPro Telemetry Server desktop is displayed, double-click My
Computer
4.
Right-click the disk drive you want to check and select Properties.
5.
Select the Tools tab.
6.
Under Error-checking, click Check Now.
7.
Check both selections and click Start.
8.
Click Yes in the Checking Local Disk window.
9.
Select Start > Shut Down >Restart, then click OK. The disk check is performed
as the ApexPro Telemetry Server reboots.
10. Repeat this procedure for the other drive.
Determine the Unity network devices that are communicating over the network by
using the Webmin Unity network function. See
on page 5-18.
127(
This procedure requires a VNC connection. See
127(
Type help at any time to see a list of commands.3
1.
Select Start > Run
2.
Type cmd, then press Enter.
3.
A command window opens displaying the prompt:
C:\Documents and settings\cic>
4.
Type cd <space> ..\..\program files\marquette\pts\3.X (X= software version)
and press Enter.
5.
Verify the prompt:
C:\Program Files\marquette\PTS\X.X> (X= software version)
6.
In the command window at the PTS\X.X prompt, type ptsconfig and press
Enter.
7.
Type display receiver and press Enter.
8.
Note the number of unassigned receivers and subtract this number from the
number of admitted beds noted above.
ApexPro Telemetry Server
2001989-178
"ApexPro Telemetry Server Service
"Unity Network Device Detection"
"VNC Logon"
on page 2-20.