Error Messages
- 2 1 4
Trigger deadlock
Indicates that the trigger source for the initiation of a measurement
is set to GET and subsequent measurement query is received. The
measurement cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.
- 2 1 5
Arm deadlock
Indicates that the arm source for the initiation of a measurement is
set to GET and subsequent measurement query is received. The
measurement cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.
-220
Parameter error
Indicates that a program data element related error occurred. This
error message 0 should be used when the device cannot detect the
more specific errors described for errors -221 through -229.
-221
Settings conflict
Indicates that a legal program data element was parsed but could
not be executed due to the current device state (see IEEE 488.2,
6.4.5.3 and 11.5.1.1.5.).
-222
Data out of range
Indicates that a legal program data element was parsed but could
not be executed because the interpreted value was outside the
legal range as defined by the device (see IEEE 488.2, 11.5.1.1.5.).
-223
Too much data
Indicates that a legal program data element of block, expression,
or string type was received that contained more data than the
device could handle due to memory or related device-specific
requirements.
-224
Illegal parameter value
Used where exact value, from a list of possibilities, was expected.
-225
Out of memory.
The device has insufkient memory to perform the requested
operation.
4-16