Advertisement

5.6 Turning OFF the Device

To turn OFF the mixer, press and hold the Control Knob (K) for 4
seconds. During this time the green power LED will start fl ashing
at a high frequency. After 4 seconds the mixer will turn OFF. At the
same time the Oxygen Valve will open to deliver a 40 L/min fl own
rate for 1 second to clear the gas pathways of residual gas.
If the Control Knob (K) is pressed and held for less than 4 seconds and
then released, the mixer will stay ON.
Note: It is important that the device is turned OFF after each patient
use and turned ON prior to use to allow the self-diagnostic protocol
to take place.

6. POST USE

6.1 Disconnect device after use
A.
Turn OFF the gas supplies to the mixer.
B.
Disconnect the gas supply hoses.
C.
Disconnect the patient circuit from the output connector and the
pressure sensor connection.
D.
Unplug the power cable from the mains supply if no charging is
required.
E.
Clean and disinfect the device and replace the single use circuit in
accordance with section 8.1 in this manual.
Note: It is important that the device is turned OFF after each patient
use and turned ON prior to use to allow the self-diagnostic protocol
to take place.

6.2 Storage

Store the mixer within the temperature and humidity ranges specifi ed
in Chapter 9.1.
Note: The mixer operates within 5 minutes after being brought back
from minimum storage temperature to room temperature;
The mixer operates within 2 minutes after being brought back from
maximum storage temperature to room temperature.
30
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)