10. TROUBLESHOOTING

Please contact the manufacturer if a problem cannot be rectifi ed. For
the safety of the patient and the health care providers DO NOT continue
Message/fault
High inhalation resistance
External leak
Extremely high
inhalation resistance
No input gas
Low input gas
Battery discharges
quickly
eAdvantage® cannot
be switched ON
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
WARNING
using the mixer.
Cause
Disconnected sensing hose
or no input gas
Input hoses are not tightened
Device is OFF
No gas or input gas pressure
below 20 PSI
Low gas below 40 PSI
No proper charging/faulty
battery
Battery empty/no power
supply connected/defective
Remedy
Check input gas valves or
sensing hose connection
Tighten input hoses
Remove mask from face and
ensure the device is ON
Change the gas cylinder
Change the gas cylinder
Charge battery as per
instructions/replace battery
Change battery/connect
power supply/ send to
O-Two™ for repair or service
43