Download Print this page

O-Two 01FC Quick Start Manual

Flow controllers

Advertisement

Quick Links

WARRANTY:
The O-Two Therapy Flow Controllers for Use with Medical Gases are manufactured from the
fi nest quality materials. Each individual part is subject to strict quality control tests to ensure
exceptionally high standards. The manufacturer warrants to the purchaser of the O-Two
Flow Controller that its component parts are free from defects in material and workmanship
for a period of fi ve (5) years from the date of purchase. The manufacturer will replace and/
or repair all parts of the Flow Controller at its option for fi ve years from the date of purchase
at no cost to the purchaser, upon the notifi cation of the defects, in writing by the purchaser
and confi rmation of said defect by the manufacturer. All shipping costs shall be borne by
the purchaser. The manufacturer shall be liable under this warranty only if the Flow Controller
and its parts have been used and serviced in the normal manner described in the instruction
manual. This warranty does not cover failure resulting from alterations, accidental damage,
misuse or operation not in accordance with the instructions provided. Wear items are not
covered hereunder. This warranty is null and void if unauthorized repairs are attempted or
performed. There are no other expressed or implied warranties. This warranty gives no
specifi c legal rights.
SPECIFICATIONS:
GAS
Medical Oxygen, Medical Air
OPERATING INLET PRESSURE
50 to 55 PSI
THERAPY FLOWRATE RANGE
0 – 6 l/min, 0 – 8 l/min, 0 – 15 l/min, 0 – 25 l/min
STORAGE TEMPERATURE
From -28°C to +60°C
OPERATING TEMPERATURE
From -18°C to +50°C
O-TWO MEDICAL TECHNOLOGIES INC.
45A Armthorpe Road, Brampton, Ontario, Canada L6T 5M4
Tel. (905) 792-6896 Fax (905) 799-1339
www.otwo.com
FLOW CONTROLLERS
FOR USE WITH MEDICAL GASES
REF
INTENDED USE:
The O-Two Medical Technologies "Therapy Flow Controller" is intended for the delivery of adjustable
therapy fl ow rates (medical oxygen or medical air) to a patient.
DESCRIPTION:
The O-Two "Therapy Flow Controller" is a therapy fl ow administration device which is capable of
delivering a controllable continuous fl ow of medical gas (oxygen or air) within the specifi ed range.
Connected to medical gas supply system, O-Two "Therapy Flow Controller" can deliver the desired
therapy fl ow by adjusting fl ow-selecting control.
1. It is imperative that the O-Two Therapy Flow Controller only be operated in accordance with the
instructions as described in this manual by clinically trained and authorized personnel. Do not use the
Flow Controller for driving any medical equipment.
2. Using oil, grease or other petroleum-based substances with oxygen and oxygen mixtures may cause
fi re or explosion. Never handle Flow Controllers with oily or greasy hands or gloves. Do not use or store
oxygen equipment near excessive heat or open fl ame.
3. Make sure that the inlet fi ttings on fl ow controllers are properly mated for the gas intended. Never
attempt to force an incompatible connection or change the connectors. Never interchange fl ow
controllers, hoses, or other equipment with similar equipment intended for use with other gases.
4. Disconnect the fl ow controller from a pressure source and release all gas from the fl ow controller
after each use.
5. Do not stand in front of a fl ow controller outlet when opening the pressure source in case foreign
particles are present which could cause a hazardous malfunction of the fl ow controller. Never occlude
the outlet to test for the presence of pressure.
6. The fl ow controller doesn't indicate that a fl ow is occurring.
7. Setting the fl ow-selecting knob between adjacent settings may cause the output fl ow to cease!
8. Varying inlet pressure outside the specifi ed operating inlet pressure has signifi cant impact on the
accuracy of output therapy fl ow rate!
9. Presence of outlet resistance may result in decrease of the fl ow rates delivered.
10. Ambient temperature variations within the specifi ed operating temperature range have no e ect
on the accuracy of fl ow rate.
11. Oxygen therapy may be critical treatment. The application of oxygen therapy using this fl ow
controller should be made in strict accordance with the prescription and instructions of a physician.
12. Never disassemble a fl ow controller under pressure! Disassembly and assembly of fl ow controllers shall
be performed only by O-Two authorized and trained personnel. Unauthorized disassembly will void the
warranty. Lubricant, if used in the maintenance of the fl ow controller, shall be compatible with the intended
medical gases. Use of lubricants not recommended by the manufacturer may cause fi re or explosion.
Rev. 7 Feb 2020
01FC
SAFETY WARNINGS

Advertisement

loading

Summary of Contents for O-Two 01FC

  • Page 1 fl ow by adjusting fl ow-selecting control. SAFETY WARNINGS 1. It is imperative that the O-Two Therapy Flow Controller only be operated in accordance with the instructions as described in this manual by clinically trained and authorized personnel. Do not use the Flow Controller for driving any medical equipment.
  • Page 2 12- AGA Probe 13- DISS Inlet- Hand tight O-Two Medical recommends that the fl ow controller be returned to the manufacturer or its authorized 14- DISS Inlet- Hand tight (US version) agents for evaluation for damage, contamination, wear and function every 5 years. The maximum service life for the fl...