Electromagnetic Compatibility - O-Two eAdvantage 01EQ3000 User Manual

Electronic nitrous oxide/oxygen analgesic gas mixing and delivering system
Hide thumbs Also See for eAdvantage 01EQ3000:
Table of Contents

Advertisement

AC/DC Power Supply
MODEL
INPUT VOLTAGE
INPUT CURRENT
INPUT FREQUENCY
OUTPUT VOLTAGE
OUTPUT CURRENT
DC OUTPUT PLUG
DC OUTPUT CABLE LENGTH
AC POWER CORD LENGTH
WEIGHT
Note: Upon disconnecting AC Power supply, Mixer will automatically
switch to battery operation without a ecting mixer behavior.

9.4 Electromagnetic Compatibility

The eAdvantage® has been tested and complies with IEC 60601-1-2:2014
requirements and is intended for use in hospital environment except near
active HF surgical equipment where the intensity of electromagnetic
disturbance is high. The user of eAdvantage® should ensure that it is not
used in environments outside those specifi ed below:
Electromagnetic Emissions
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC61000-3-2
Voltage fl uctuations/
fl icker emissions
IEC61000-3-3
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)
PMP90-13-2 01CV0104
100 - 240 VAC
1.06 - 0.45 A
47 - 63 Hz
19 VDC
4.74 A Maximum
2.5 x 5.5 x 11 mm
6'
6'
642 g
Compliance
Electromagnetic environment - guidance
The eAdvantage® uses RF energy only for its
internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
The eAdvantage® external power supply is
Class A
suitable for use in all establishments other than
domestic and those directlyly connected to the
public Low-voltage power supply network that
supplies buildings used for domestic purposes.
Class A
If it is used in a residential environment,
the eAdvantage® might not o er adequate
protection to RF communication services. The
user might need to take mitigation measures,
such as relocation or re-orienting the device.
Complies
41