Create Declaration of Conformance from T-Box XL™ Conformance Data
Note: Declarations of Conformance can only be produced for ISO 6789 tool types.
Declarations of Conformance do not use uncertainty values in the respective tool templates 'Uncertainty' tab.
1.
Conformance test a Tool on T-Box XL™
and then Synchronize with TDMS.
2.
Select Conformance Data to show results
in the right hand pane.
3.
Right click on the Conformance Data to show
the drop down menu.
4.
Select Create Conformance Declaration
5. Select or enter an Inspector
6. Enter Certificate Number (if required)
7. Enter the Temperature (in degrees C) that the calibration was performed
8. Enter Certificate Text
Choose either As Found or As Left from the drop down
or enter your own text.
9.
If the same tool has other conformance data to put on the same declaration of conformance, press
Select Results
This feature allows clockwise & anti-clockwise as well as 'as found' & 'as left' results to be included on
the same declaration of conformance.
To add another calibration to the same declaration of conformance double click on the required results.
60
(this must be
entered).
(this must be
entered).