TRANSDUCERS
Any transducers used for ISO 6789:2017 Calibrations or Conformances or non-ISO Calibrations should be
added to the Transducer register.
Note: If any of the transducers are not SMART, entering either of the mV/V values will enable them to
synchronise with the T-Box XL™ instrument.
Add Transducer
1.
Select Define menu then select Transducers
(or select
).
Transducers are shown in the right hand pane.
2.
Right click on a Transducer to show
the drop down menu.
3.
Select Add Transducer.
SMART transducers:-
Fill in:-
Serial Number
Model
Calibration Certificate No./s
Capacity
Units
Transducer Uncertainty
Transducer Uncertainty Interval
The 2 uncertainty values should be the highest
'Expanded Uncertainty' and 'Uncertainty
Interval' values from the certificate/s
All other fields should be left blank
Non-SMART transducers:-
Fill in:-
Serial Number
Model
Calibration Certificate No./s
Capacity
Units
Transducer Uncertainty
Transducer Uncertainty Interval
mV/V value/s
Angle Measurement data (if applicable)
Note: Only Non-Smart transducers will be
synchronised with a T-Box XL™
40