8. Technical Data

General Data
Information
AC Adaptor
Sartorius AC adaptor module
Primary
Secondary
Further data
Balance
Power supply
Input voltage
Power consumption
Further data
Ambient Conditions
The specifications apply under the following ambient conditions:
Environment
Ambient temperature*
Operational capability
Storage and shipping
Elevation
Relative humidity**
Safety of Electrical Equipment
Electromagnetic Compatibility
Defined immunity to interference
Interference emission
Standard Equipment
1
Selectable weight units
Available application programs
Verified balances in accordance with EU requirements comply with the requirements of the Council Directive 2009/23/EG with
EN 45501:1992 and OIML R76:2006.
*
For verified balances in accordance with EU requirements, refer to the information on the balance.w
**
For verified balances in accordance with EU requirements, the legal regulations apply.
1
Depending on the country-specific model version, not all weight units listed may be available.
Unit
Value
100 –240 V~, –10% / +10%, 50–60 Hz, 0.2 A
15 V DC, ± 5%, 530 mA (max.) / 8 W (max.): 0 to +40 °C
15 V DC, ± 5%, 330 mA (max.) / 5 W (max.): 0 to +50 °C
Protection class II in accordance with EN / IEC 60950-1 | up to 3000 m above
sea level
IP40 in accordance with EN 60529 / IEC 60529
Only via Sartorius AC adaptor module 6971790
V DC
12.0 – 15.0
W
4.5 (typically)
IP43 in accordance with EN 60529 / IEC 60529
For indoor use only
°C
+10 – +30
°C
Guaranteed between +5 – +45
°C
-10 – +60
m above
Up to 3000
sea level
%
15 – 80% for temperatures up to 31 °C non-condensing, decreasing linearly
up to 50% relative humidity at 40 °C and 20% at 50 °C.
As per EN 61010-1 / IEC61010-1 Safety requirements for electrical equipment
for measurement, control and laboratory use – Part 1: General requirements
As per EN 61326-1 / IEC61326-1 Electrical equipment for measurement,
control and laboratory use – EMC requirements – Part 1: General requirements
Suitable for use in industrial areas
Class B
Suitable for use in residential areas and areas that are connected to a low
voltage network that also supplies residential buildings.
Gram, kilogram, carat, pound, ounce, troy ounce, Hong Kong teal, Singapore
teal, Taiwan teal, grain, pennyweight, milligram, parts per pound, China teal,
mommas, Austrian carat, tola, baht, mesghal and Newton
®
Secura
Models
Counting, weighing in percent, density determination, peak hold, unstable
condition, checkweighing, mixing, components (totalizing), statistics,
conversion, SQmin and identifiers
®
Quintix
Models
Counting, weighing in percent, density determination, peak hold, unstable
condition, checkweighing, mixing, components (totalizing), statistics,
conversion
English
(Model-specific data see page 80)
Installation Instructions Secura, Quintix
9